Product
Specification
|
Description
|
|
Conductivity Range
|
0.00 to 19.00 mΩ/cm
0.000 to 1.999 mΩ/cm
|
0.01 mΩ/cm
0.001 mΩ/cm
|
|
Accuracy
|
±1 % full scale
|
|
Temperature Range
|
-9.9 to 125 °C
|
|
Resolution &
Relative Accuracy
|
0.1 °C &
±0.5 °C
|
|
Temperature Sensor
|
Pt 100/ Pt 1000 (jumper selectable)
|
|
Temperature Compensation
|
Auto / Manual (reference at 25
°C)
|
|
Set Point
and Controller Functions
|
|
Function
|
Limit / Proportional (pulse or
frequency) controller
|
|
Switching Hysteresis
|
0 to 10% full scale
|
|
Pickup / Dropout Delay
|
0 to 2000 seconds
|
|
Contact Outputs, Controller
|
2 potential free change-over contacts
|
|
Switching Voltage / Current /
Power
|
Max 250 VAC / Max 3A / Max 600VA
|
|
Alarm
Functions
|
|
Function (Switchable)
|
Launching or Pulse
|
|
Pickup Delay
|
0 to 2000 seconds
|
|
Switching Voltage / Current /
Power
|
Max 250 VAC / Max 3A / Max 600VA
|
|
Electrical
Data and Connections
|
|
Power requirement
|
110/ 220 VAC (jumper selection)
|
|
Frequency
|
48 to 62Hz
|
|
Signal Output
|
0/4 to 20mA, galvanically isolated
|
|
Load
|
Max. 600 Ω
|
|
Connection Terminal
|
Terminal blocks 5-pole/ 17-pole,
removable
|
|
Main Fuse/ Fine Wire Fuse
|
Slow-blow 250 V/ 100 mA
|
|
EMC Specifications
|
|
Emitted Interference
|
According to EN 50081-1
|
|
Immunity to Interference
|
According to EN 50082-1
|
|
Environmental
Conditions
|
|
Maximum Relative Humidity
|
10 to 95%, non-condensing
|
|
Ambient Temp. Operating Range
|
0 to 50 °C
|
|
Mechanical
Specifications
|
|
Dimensions
|
175 x 96 x 96 mm
|
|
Weights
|
Max. 0.7kg
|
|
Material
|
ABS with polycarbonate
|
|
Insulation
|
IP 54/ IP 65
|