Product Specification
|
Description
|
|
Conductivity range
|
*(R1): 0-2 µS/cm
R1 (R2): 0-20 µS/cm
R2 (R3): 0-200 µS/cm
R3 (R4): 0-2000 µS/cm
R4 (R5): 0-20 mS/cm
R5 (R6): 0-200 mS/cm
|
|
Resolution
|
0.001 µS/cm
0.01 µS/cm
0.1 µS/cm
1 µS/cm
0.01 mS/cm
0.1 mS/cm
|
|
Accuracy
|
±1% of full scale reading in respective
range
|
|
Temperature
|
-10 °C to 125 °C
(14 °F to 257 °F)
|
|
Resolution
|
0.1 °C
(0.1 °F)
|
|
Relative Accuracy
|
±0.5 °C
(±0.9 °F)
|
|
Sensor
|
Pt100 / Pt1000 (3 wire / 2 wire)
(selectable)
|
|
Temperature Compensation
|
Auto/Manual (reference at 25 °C)
|
|
Cell Constant
|
User-selectable (0.10 to 10.00)
|
|
Temperature Coefficient
|
0 to 10% (selectable)
|
|
Normalisation Temperature
|
25 °C
|
|
Controller Function
|
Limit Control
|
|
Relay Operating Parameters
|
LO (Relay active when measured
value < set point)
HI (Relay active when measured value > set
point
|
|
Time Delay
|
0 to 2000 seconds
|
|
Switching Conductivity Hysteresis
|
0 to 10% of full scale
|
|
Contact Output
|
2 SPST relay
|
|
Switching Voltage/Current Power
|
Max 250VAC/Max 3A/MAX 600 VA
|
|
Display
|
Dual Display Custom LCD
|
|
Backlight
|
On/Off selectable
|
|
Power Requirements
|
110 or 220 VAC to 9 V DC Adapter
|
|
Dimension
|
96 x 96 x 66 mm (3.78 x 3.78 x
2.6 inch)
|
|
Weight
|
210 g
|