Industrial Process Controllers & Monitors
alpha-pH800 1/4 DIN pH/ORP Controller
User-Customization through Advanced Setup Menu offers flexibility in matching the controller's functions to suit individual's specific requirement
Automatic Calibration with Auto-Buffer Recognition eliminates mistakes during calibration
Symmetrical Mode Operation eliminates electronic noise problems when used with solution ground
One-Point Online Calibration without shutting down the line
Hold Relay for use with float switches/flow switches and other controllers as a failsafe function
0 to 2000 Second Time Delay Adjustment on control and alarm delays
Galvanically Isolated Scaleable 0-20/4-20 mA Outputs for pH/ORP
Hysteresis Band Protection against rapid control switching of relays from wear and tear
Non-Volatile Memory retains all stored parameters and calibration data even if power fails
Large Dual Display shows pH (or ORP) with temperature simultaneously - features clear multiple icons, set points, and status messages
Choice of Temperature Sensor Pt100/Pt1000 with 2-wire or 3-wire temperature input selection
Easy Installation and Wiring with detachable plug-in connectors
Applications:
General: Useful for any batch or on-line type application that requires accurate pH or ORP control.
Water Purification/Treatment: Use for batch and on-line control of incoming process water, rinse water treatment, recirculating system and waste water treatment.
Industrial: Ideal for chemical processing, food processing. aquarium, pharmaceutical, hydroponics and waste control industries.
Specifications:
|
Product
Specification
|
Description
|
|
pH Range
|
0.00 to 14.00 pH
|
|
Resolution & Accuracy
|
0.01 pH & ±0.01 pH
|
|
mV Range
|
-999 to 1000 mV
|
|
Resolution & Accuracy
|
1/ ±2 mV
|
|
Temperature Range
|
-10 to 110 °C
|
|
Resolution & Relative Accuracy
|
0.1 °C & ±0.5 °C
|
|
Temperature Sensor
|
Pt 100 (3-wire)
|
|
Temperature Compensation
|
Auto / Manual (0°C to 100°C)
|
|
Set Point
and Controller Functions
|
|
Function
|
Limit Control
|
|
Switching pH/ORP Hysteresis
|
0.00 to 2.00 pH or 0 to 100 mV
|
|
Contact Outputs, Controller
|
2 SPDT relays
|
|
Switching Voltage/ Current/ Power
|
Max 250 VAC/ Max 3A/ Max 600 VA
|
|
Electrical
Data and Connections
|
|
Power Requirement
|
85 to 260 VAC, 50/60 Hz or DC
|
|
Signal Output/ Load
|
4-20mA galvanically isolated /
200Ω
|
|
pH/ORP Input
|
BNC (1012 Impedence)
|
|
Connection Terminals
|
3 detachable connectors (3-pin;
7-pin & 9-pin terminal blocks)
|
|
Main Fuse
|
250mA, Anti-surge (BUSSMAN S504
+ 250mA)
|
|
Environmental
Conditions
|
|
Maximum Relative Humidity
|
80% up to 31°C decreasing linearly
to 50% at 40°C
|
|
Ambient Temp. Operating Range
|
-10 to 50 °C (14 °F to 122°F)
|
|
Mechanical
Specifications
|
|
Dimensions
|
96 x 48 x 150 mm
|
|
Weights
|
300 g (boxed)
|
|